Skip to main content

Table 1 Major blood pressure device validation protocols

From: History and evolution of blood pressure measurement

Protocol

Year published

Sample size required (No. of paired readings)

Criteria indicating a valid device

BHS [48]

1990

85 (255)

Device graded from A to D. Grade A is the highest level of accuracy and requires that the percentage of readings with a difference between the device-under-test and the reference sphygmomanometer of ≤5, 10, and 15 mm Hg be 65, 85, and 95%, respectively.

ESH International Protocol [49]

2010

33 (99)

Pass requirements are split into 2 phases and are based on the number of measurements with differences between the device-under-test and reference sphygmomanometer of ≤5, 10, and 15 mm Hg. See protocol for details. This protocol is being phased out, to be replaced by a joint universal AAMI/ESH/ISO validation protocol requiring 85 subjects.

AAMI/ANSI/ISO [50]

2013

85 (255)

• Criterion 1: when analyzed as 255 paired determinations, the mean difference between the device-under-test and reference sphygmomanometer is < 5.0 mm Hg, and the SD of the difference is < 8.0 mm Hg.

• Criterion 2: when analyzed as 85 paired determinations, the SD of the difference between the device-under-test and reference sphygmomanometer is < 4.79 to 6.95 mm Hg (the actual threshold varies according to the mean difference observed. See protocol for details).

AAMI/ESH/ISO [51]

2018

85

• A device is considered acceptable if its estimated probability of a tolerable error (≤10 mmHg) is at least 85%.

• The mean BP difference (test versus reference) and its SD, criteria 1 and 2 of the ANSI/AAMI/ISO 81060–2, to be applied for systolic and diastolic BP.

  1. AAMI indicates Association for the Advancement of Medical Instrumentation, ANSI American National Standards Institute, BHS British Hypertension Society, ESH European Society of Hypertension, and ISO International Organization for Standardization